How Well Does AI Classify Bone Marrow Cells? Evaluating the MC-06 Across 50 Smears and 25,000 Nucleated Cells

MC-06 Cell Morphology Analyzer

Software: Blood Cell Analysis System

Commissioning organization: Zhongshan Xinrui Medical Equipment Technology Co., Ltd.

Clinical evaluation site:The First Hospital of China Medical University

Completion date:1 November 2026

Report number: XR-MC06-PE-2026-01

Version: V1.0

Abstract

Objective. To evaluate the performance and clinical utility of the MC-06 and its Blood Cell Analysis System for automated differential counting of nucleated cells in bone marrow smears.

Methods. Fifty bone marrow smears covering common hematologic disorders and approximately normal marrow were selected at the clinical evaluation site. For each smear, 500 nucleated cells were counted by each of three methods: AI preclassification, AI preclassification followed by expert review, and manual microscopy. Expert-reviewed classifications provided the reference for the confusion matrix, sensitivity, specificity, accuracy and Cohen’s kappa. Manual microscopy provided the reference for Pearson correlation and Wilcoxon signed-rank analyses of the unreviewed and reviewed results.

Results. Mean sensitivity (recall) was 95.71%, mean specificity 99.73%, overall accuracy (agreement) 95.71%, mean false-positive rate 0.27%, and mean false-negative rate 4.29%. Overall kappa was 0.9521, with κ ≥ 0.80 in 14 of 16 categories. AI preclassification correlated positively with manual microscopy for all 16 categories (P < 0.05), with r ≥ 0.85 for 10. After review, systematic differences for orthochromatic erythroblasts, lymphocytes and plasma cells were no longer statistically significant. Morphologic findings were concordant with the clinical diagnosis in 49/50 cases (98%).

Conclusion. The analyzer performed well in marrow nucleated cell differential counting, with high agreement with expert review and strong correlations with manual microscopy. Expert review further corrected residual AI preclassification bias. The system operated reliably and provided traceable results, supporting its use as an aid in clinical laboratories.

Keywords: cell morphology analyzer, artificial intelligence, bone marrow cytomorphology, AI preclassification, expert review, manual microscopy, performance evaluation.

Background and objectives

Background

Bone marrow cytomorphology is fundamental to the diagnosis, classification, treatment response assessment and prognostic evaluation of hematologic disorders. Conventional microscopy requires a laboratory physician to classify 200–500 nucleated cells field by field. This is labor-intensive and time-consuming, and interpretations vary between readers. The assessment process is difficult to document and review in a standardized manner. Advances in digital imaging and deep learning have enabled integrated smear scanning, automated cell detection and classification, and structured reporting.

The MC-06, developed by the commissioning organization, acquires high-quality marrow cell images by automated scanning. AI algorithms detect, classify and count nucleated cells, providing objective, reproducible and traceable quantitative information. This study evaluated analytical performance and clinical usability.

Study objectives

Using three paired differential counts from each of 50 smears, the study assessed (1) agreement between AI preclassification and expert review, (2) category-specific correlations between AI preclassification and manual microscopy, (3) correlations after expert review and the effect of review on systematic bias, and (4) concordance between AI-derived morphologic findings and the clinical diagnosis.

Device workflow and features

Workflow

The analyzer scans the entire Wright–Giemsa-stained smear with a 10× objective and identifies an area near the junction of the body and feathered edge with evenly distributed cells and no obvious compression. A 100× oil-immersion objective then captures high-resolution images. After denoising and enhancement, a detection network locates and crops individual cells. A classification network assigns each cell to a category. A laboratory physician reviews and corrects the classifications at the workstation. The system produces a structured report with differential counts, percentages and archived images.

Article content
MC-06 WORKFLOW

Features

Automated scanning, recognition, counting and reporting reduce manual steps. AI preclassification followed by physician review provides two levels of quality control. Linked images and classifications enable retrospective review and quality tracing. Uniform classification and interpretation criteria reduce inter observer variability.

Procedure

1. Register specimens. Assign a unique study identifier to each adequately stained, acceptable smear and record specimen information, patient characteristics and clinical diagnosis.

2. Perform AI preclassification. Scan all 50 slides in bone marrow mode according to the operating procedure. Following whole-slide scanning at 10× and high-power acquisition at 100×, export the automated 500-cell differential report for each specimen.

3. Review classifications. A laboratory physician qualified in diagnostic marrow morphology reviews each cell on the workstation, corrects classifications where necessary and exports the reviewed differential.

4. Perform manual microscopy. A senior laboratory physician independently classifies and records 500 nucleated cells per specimen according to WHO classification criteria and Chinese expert consensus recommendations.

5. Verify data. Enter and cross-check all three datasets using uniform criteria before statistical analysis.

Statistical analysis

SPSS 26.0 was used. A confusion matrix comparing AI preclassification with expert review yielded true-positive, false-positive, false-negative and true-negative counts for each of 16 categories. Sensitivity, specificity, accuracy and false-positive and false-negative rates were calculated. Cohen’s kappa assessed agreement. Pearson correlation assessed associations with manual microscopy. Scatterplots show manual microscopy on the x-axis and AI preclassification on the y-axis, with fitted linear regressions. The Wilcoxon signed-rank test assessed systematic differences. P < 0.05 was considered statistically significant.

Results

Visualization and traceability

Across 50 specimens, the system acquired 14,234 high-resolution 100× fields from selected regions and detected, cropped and classified 42,903 individual nucleated cells. Nuclear structure, chromatin, cytoplasm and granules were clearly visible. Each cell image was archived with its classification for review and traceability.

Article content
Representative cropped images of nucleated bone marrow cells.
Article content
Representative cropped images of nucleated bone marrow cells.

Conclusions

The evaluation of the MC-06 and its Blood Cell Analysis System at The First Hospital of China Medical University reached the following conclusions.

1. Classification accuracy was good. Against expert review, mean sensitivity was 95.71%, mean specificity 99.73%, overall agreement 95.71%, mean false-positive rate 0.27% and mean false-negative rate 4.29%. The primary measures exceeded the 90% target. Overall Cohen’s κ of 0.9521 indicated high agreement.

2. AI preclassification correlated positively and significantly with manual microscopy for all 16 categories, with r ≥ 0.85 in 10. Systematic differences were within approximately 1–2 percentage points.

3. Expert review improved agreement with manual microscopy. The report describes category-specific correlations as improved or maintained, and systematic differences in several categories became nonsignificant after review.

4. Morphologic concordance with the clinical diagnosis was 98%. AI preclassification can assist laboratory physicians with morphologic screening and flag findings suggestive of hematologic disorders.

5. All functional modules operated reliably, with no failures during testing.

6. Linked slide images, individual cell crops, AI classifications and review records ensure traceability and allow retrospective verification, supporting routine clinical use.

Taken together, the findings indicate good recognition of nucleated marrow cells, high agreement with expert review and strong correlations with manual microscopy. The investigators concluded that the system provides accurate, stable and traceable results, can reduce workload while maintaining quality, and can improve efficiency and standardization as an adjunct to bone marrow cytomorphology in clinical laboratories.

#Hematology #BoneMarrowMorphology #DigitalMorphology #LaboratoryMedicine #ClinicalPerformanceEvaluation